The Spectrum Application (RF Measurements)
R&S
®
FPL1000
329User Manual 1178.3370.02 ─ 03
Display a certain number of peaks per range
Display only peaks that exceed a threshold ( "Margin" )
Display detected peaks as blue squares in the diagram, as well as in the peak list
Furthermore, you can save the evaluation list to a file.
Retrieving Results via Remote Control
The measured spurious values of the displayed trace can be retrieved using the
TRAC:DATA? SPUR command (see TRACe<n>[:DATA] on page 703).
7.2.7.3 Spurious Emissions Basics
Some background knowledge on basic terms and principles used in Spurious Emis-
sions measurements is provided here for a better understanding of the required config-
uration settings.
Ranges and Range Settings................................................................................. 329
Limit Lines in Spurious Measurements................................................................. 330
Ranges and Range Settings
Conditions for ranges
The following rules apply to ranges:
The minimum span of a range is 20 Hz.
The individual ranges must not overlap (but can have gaps).
The maximum number of ranges is 30
The maximum number of sweep points in all ranges is limited to 100001.
You can define a span that is smaller than the combined span of the ranges. In this
case, the measurement includes only the ranges that lie within the defined span and
have a minimum span of 20 Hz.
Measurements and Results