User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
342User Manual 1178.3370.02 ─ 03
7.2.8.4 Time Domain Power Configuration
Access: "Overview" > "Select Measurement" > "Time Domain Power" > "Time Dom
Power Config"
The remote commands required to perform these tasks are described in Chapter 8.6.8,
"Measuring the Time Domain Power", on page 602.
Results ....................................................................................................................... 342
Limit State .................................................................................................................. 342
Left Limit / Right Limit .................................................................................................342
Results
Activates the power results to be evaluated from the displayed trace or a limited area
of the trace.
"Peak"
Peak power over several measurements (uses trace averaging, Max
Hold)
"RMS"
RMS value from the points of the displayed trace or a segment
thereof.
"Mean"
Mean value from the points of the displayed trace or a segment
thereof. The linear mean value of the equivalent voltages is calcula-
ted.
"Std Dev"
The standard deviation of the measurement points from the mean
value.
The measurement of the mean power is automatically switched on at
the same time.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe] on page 605
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak:RESult? on page 608
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe] on page 605
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult? on page 609
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe] on page 604
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult? on page 607
Limit State
Switches the limitation of the evaluation range on or off. Default setting is off.
If deactivated, the entire sweep time is evaluated. If switched on, the evaluation range
is defined by the left and right limit. If only one limit is set, it corresponds to the left limit
and the right limit is defined by the stop frequency. If the second limit is also set, it
defines the right limit.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] on page 717
Left Limit / Right Limit
Defines a power level limit for line S1 (left) or S2 (right).
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT on page 717
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt on page 717
Measurements and Results