User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
412User Manual 1178.3370.02 ─ 03
Remote command:
TRIGger[:SEQuence]:LEVel:IFPower on page 664
TRIGger[:SEQuence]:LEVel:IQPower on page 665
TRIGger[:SEQuence]:LEVel[:EXTernal<port>] on page 664
Repetition Interval
Defines the repetition interval for a time trigger. The shortest interval is 2 ms.
The repetition interval should be set to the exact pulse period, burst length, frame
length or other repetitive signal characteristic.
Remote command:
TRIGger[:SEQuence]:TIME:RINTerval on page 666
Drop-Out Time
Defines the time the input signal must stay below the trigger level before triggering
again.
For more information on the drop-out time, see "Trigger Drop-Out Time" on page 407.
Remote command:
TRIGger[:SEQuence]:DTIMe on page 663
Trigger Offset
Defines the time offset between the trigger event and the start of the sweep.
For more information, see "Trigger Offset" on page 406.
Offset > 0: Start of the sweep is delayed
Offset < 0: Sweep starts earlier (pretrigger)
Only possible for zero span (e.g. I/Q Analyzer application) and gated trigger switched off
Maximum allowed range limited by the sweep time:
Pretrigger
max
= sweep time
max
For the "Time" trigger source, this function is not available.
Remote command:
TRIGger[:SEQuence]:HOLDoff[:TIME] on page 663
Hysteresis
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
This setting is only available for "IF Power" trigger sources. The range of the value is
between 3 dB and 50 dB with a step width of 1 dB.
For more information, see "Trigger Hysteresis" on page 406.
Remote command:
TRIGger[:SEQuence]:IFPower:HYSTeresis on page 664
Trigger Holdoff
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Trigger and Gate Configuration