User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
426User Manual 1178.3370.02 ─ 03
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 713
CALCulate<n>:DELTamarker<m>[:STATe] on page 711
Marker Position X-value
Defines the position (x-value) of the marker in the diagram. For normal markers, the
absolute position is indicated. For delta markers, the position relative to the reference
marker is provided.
Remote command:
CALCulate<n>:MARKer<m>:X on page 714
CALCulate<n>:DELTamarker<m>:X on page 712
Frame (Spectrogram only)
Spectrogram frame the marker is assigned to.
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe on page 729
CALCulate<n>:DELTamarker<m>:SPECtrogram:FRAMe on page 734
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal" , the type for delta marker 1 is always
"Delta" . These types cannot be changed.
Note: If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
A normal marker indicates the absolute value at the defined position
in the diagram.
"Delta"
A delta marker defines the value of the marker relative to the speci-
fied reference marker (marker 1 by default).
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 713
CALCulate<n>:DELTamarker<m>[:STATe] on page 711
Reference Marker
Defines a marker as the reference marker which is used to determine relative analysis
results (delta marker values).
If the reference marker is deactivated, the delta marker referring to it is also deactiva-
ted.
If a fixed reference point is configured (see "Defining a Fixed Reference"
on page 429), the reference point ( "FXD" ) can also be selected instead of another
marker.
Remote command:
CALCulate<n>:DELTamarker<m>:MREF on page 711
Marker Usage