User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
434User Manual 1178.3370.02 ─ 03
"Up"
Determines the next maximum/minimum above the current peak (in
more recent frames).
"Absolute"
Determines the next maximum/minimum above or below the current
peak (in all frames).
"Down"
Determines the next maximum/minimum below the current peak (in
older frames).
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:ABOVe on page 730
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:ABOVe
on page 735
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:BELow on page 731
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:BELow
on page 736
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT on page 731
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:NEXT on page 736
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:ABOVe on page 732
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:ABOVe
on page 737
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:BELow on page 732
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:BELow
on page 737
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:NEXT on page 732
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:NEXT on page 737
Marker Search Type
Defines the type of search to be performed in the spectrogram.
"X-Search"
Searches only within the currently selected frame.
"Y-Search"
Searches within all frames but only at the current frequency position.
"XY-Search"
Searches in all frames at all positions.
Remote command:
Defined by the search function, see Chapter 8.8.8.6, "Marker Search (Spectrograms)",
on page 728
Marker Search Area
Defines which frames the search is performed in.
"Visible"
Only the visible frames are searched.
"Memory"
All frames stored in the memory are searched.
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:SARea on page 730
CALCulate<n>:DELTamarker<m>:SPECtrogram:SARea on page 734
Exclude LO
If activated, restricts the frequency range for the marker search functions.
Marker Usage