User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
450User Manual 1178.3370.02 ─ 03
Remote commands:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe] on page 748
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult? on page 748
Band Power Measurement State ............................................................................... 450
Span ...........................................................................................................................450
Power Mode ...............................................................................................................451
Switching All Band Power Measurements Off ............................................................451
Band Power Measurement State
Activates or deactivates band power measurement for the marker in the diagram.
Band power markers are only available for standard frequency measurements (not
zero span) in the Spectrum application.
If activated, the markers display the power or density measured in the band around the
current marker position.
For details see Chapter 7.9.4.6, "Measuring the Power in a Channel (Band Power
Marker)", on page 448.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe] on page 748
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe] on page 750
Span
Defines the span (band) around the marker for which the power is measured. The span
is indicated by lines in the diagram.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN on page 748
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN on page 750
Marker Usage