You can turn trace smoothing on and off for all traces individually and compare, for
example, the raw and the smooth trace.
Linear smoothing is based on the following algorithm:
2
1
2
1
1
'
n
sx
n
sx
xy
n
sy
Equation 7-1: Linear trace smoothing
With:
s = sample number
x = sample offset from s
n = aperture size
7.10.1.2 Trace Settings
Access: "Overview" > "Analysis" > "Traces"
Or: TRACE > "Trace Config"
You can configure the settings for up to 6 individual traces.
For settings on spectrograms, see Chapter 7.10.2.2, "Spectrogram Settings",
on page 479.
Trace data can also be exported to an ASCII file for further analysis. For details see
Chapter 7.13.1, "Trace / Data Export Configuration", on page 512.
Trace Configuration