518User Manual 1178.3370.02 ─ 03
File contents Description
Level Range;100;dB Display range in y direction. Unit: dB with x-axis LOG, % with x-
axis LIN
Rf Att;20;dB Input attenuation
RBW;100000;Hz Resolution bandwidth
VBW;30000;Hz Video bandwidth
SWT;0.005;s Sweep time
Trace Mode;AVERAGE; Display mode of trace: CLR/WRITE,AVER-
AGE,MAXHOLD,MINHOLD
Detector;AUTOPEAK; Selected detector
Sweep Count;20; Number of sweeps set
Data section
Trace 1:;; Selected trace
x-Unit;Hz; Unit of x values: Hz with span > 0; s with span = 0; dBm/dB
with statistics measurements
y-Unit;dBm; Unit of y values: dB*/V/A/W depending on the selected unit
with y-axis LOG or % with y-axis LIN
Values; 1001; Number of measurement points
Frames;2; Number of exported frames
Frame;0; Most recent frame number
Timestamp;17.Mar 11;11:27:05.990 Timestamp of this frame
10000;-10.3;-15.7
10130;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values, identical to spectrum data:
<x value>, <y1>, <y2>; <y2> being available only with detector
AUTOPEAK and containing in this case the smallest of the two
measured values for a measurement point.
Frame;-1; Next frame
Timestamp;17.Mar 11;11:27:05.342 Timestamp of this frame
...
7.14 Optimizing Measurements
7.14.1 Minimizing the Measurement Duration
If you want to minimize the measurement duration, try the following methods to opti-
mize the measurement:
●
Reduce the span of the measurement to the relevant parts of the signal only.
Optimizing Measurements