522User Manual 1178.3370.02 ─ 03
*SRE............................................................................................................................ 524
*STB?...........................................................................................................................524
*TRG............................................................................................................................ 525
*TST?...........................................................................................................................525
*WAI.............................................................................................................................525
*CAL?
Calibration query
Initiates a calibration of the instrument and subsequently queries the calibration status.
Responses > 0 indicate errors.
Note: If you start a self-alignment remotely, then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed.
Usage:
Query only
Manual operation: See " Start Self Alignment " on page 160
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of
the QUEStionable and the OPERation registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE <Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value> Range: 0 to 255
*ESR?
Event status read
Returns the contents of the event status register in decimal form and subsequently
sets the register to zero.
Return values:
<Contents> Range: 0 to 255
Usage: Query only