8.5.2 Performing a Sequence of Measurements
The following commands control the sequencer.
For details on the Sequencer see
INITiate<n>:SEQuencer:ABORt.......................................................................................530
INITiate<n>:SEQuencer:IMMediate..................................................................................530
INITiate<n>:SEQuencer:MODE....................................................................................... 531
SYSTem:SEQuencer......................................................................................................531
INITiate<n>:SEQuencer:ABORt
This command stops the currently active sequence of measurements. The Sequencer
itself is not deactivated, so you can start a new sequence immediately using
INITiate<n>:SEQuencer:IMMediate on page 530.
To deactivate the Sequencer use SYSTem:SEQuencer on page 531.
Suffix:
<n>
.
irrelevant
Usage: Event
Manual operation: See " Sequencer State " on page 88
INITiate<n>:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Before this command can be executed, the Sequencer must be activated (see
SYSTem:SEQuencer on page 531).
Suffix:
<n>
.
irrelevant
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Usage: Event
Manual operation: See " Sequencer State " on page 88