User Manual

Remote Commands
R&S
®
FPL1000
537User Manual 1178.3370.02 ─ 03
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Sets the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
Usage: Event
Manual operation: See " Frequency Sweep " on page 233
See " Zero Span " on page 233
See " Single Sweep / Run Single " on page 401
[SENSe:]SWEep:COUNt:CURRent?
This query returns the current number of started sweeps or measurements. This com-
mand is only available if a sweep count value is defined and the instrument is in single
sweep mode.
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage: Query only
8.6.2 Configuring Power Measurements
The following commands work for several power measurements.
CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE...................................................... 537
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?.............................................538
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect..................................................... 540
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]....................................................540
[SENSe:]POWer:ACHannel:PRESet................................................................................ 541
[SENSe:]POWer:ACHannel:PRESet:RLEVel.................................................................... 541
[SENSe:]POWer:TRACe.................................................................................................541
CALCulate<n>:MARKer<m>:FUNCtion:POWer:MODE <Mode>
This command selects the trace display mode for power measurements.
Suffix:
<n>
.
Window
<m> Marker
Configuring and Performing Measurements