Parameters:
<Standard> For more information see Chapter 7.2.3.8, "Reference: Prede-
fined CP/ACLR Standards", on page 277.
If you want to load a customized configuration, the parameter is
a string containing the file name.
Manual operation: See " Predefined Standards " on page 263
See " User Standards " on page 263
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:CATalog?
This command queries all files containing ACLR standards.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<Standards> List of standard files.
Usage: Query only
Manual operation: See " User Standards " on page 263
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:DELete <Standard>
This command deletes a file containing an ACLR standard.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Standard> String containing the file name of the standard.
Usage: Event
Manual operation: See " User Standards " on page 263
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:SAVE <Standard>
This command saves the current ACLR measurement configuration as a new ACLR
standard.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Suffix:
<n>
.
Window
<m> Marker
Configuring and Performing Measurements