Parameters:
<State> ON | OFF | 1 | 0
Relative limit check for lower adjacent channel
*RST: 0
<State> ON | OFF | 1 | 0
Relative limit check for upper adjacent channel
*RST: 0
Manual operation: See " Limit Check " on page 268
CALCulate<n>:LIMit<k>:ACPower:ALTernate<ch>:ABSolute
<LowerLimit>,
<UpperLimit>
This command defines the absolute limit of the alternate channels.
If you have defined an absolute limit as well as a relative limit, the R&S FPL1000 uses
the lower value for the limit check.
Suffix:
<n>, <k>
.
irrelevant
<ch> 1 to 11
Alternate channel number
Parameters:
<LowerLimit> The limit of the lower adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
<UpperLimit> The limit of the upper adjacent channel.
Range: -200 dBm to 200 dBm
*RST: -200 dBm
Manual operation: See " Limit Check " on page 268
CALCulate<n>:LIMit<k>:ACPower:ALTernate<ch>:ABSolute:STATe
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<k>:ACPower[:STATe].
Suffix:
<n>, <k>
.
irrelevant
<ch> 1 to 11
Alternate channel number
Manual operation: See " Limit Check " on page 268