User Manual

Remote Commands
R&S
®
FPL1000
556User Manual 1178.3370.02 ─ 03
TRACe<n>[:DATA] on page 703
Remote commands exclusive to channel power measurements
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ.............................................. 556
[SENSe:]POWer:ACHannel:MODE..................................................................................556
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ <State>
This command selects the way the R&S FPL1000 returns results for power measure-
ments.
You can query results with CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:
RESult?.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
Channel power density in dBm/Hz
OFF | 0
Channel power in dBm
*RST: 0
Example:
CALC:MARK:FUNC:POW:RES:PHZ ON
Output of results referred to the channel bandwidth.
Manual operation: See "Channel power level and density ( Power Unit )"
on page 265
[SENSe:]POWer:ACHannel:MODE <Mode>
This command selects the way the R&S FPL1000 displays the power of adjacent chan-
nels.
You need at least one adjacent channel for the command to work.
Parameters:
<Mode> ABSolute
Shows the absolute power of all channels
RELative
Shows the power of adjacent and alternate channels in relation
to the transmission channel
*RST: RELative
Manual operation: See " Absolute and Relative Values (ACLR Mode) " on page 265
Configuring and Performing Measurements