Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the measurement range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STARt -10
For a detailed example see Chapter 8.6.6.10, "Example: SEM
Measurement", on page 587.
Manual operation: See " Abs Limit Start / Stop <n> " on page 305
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:ABSolute:STOP
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the measurement range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STOP -15
For a detailed example see Chapter 8.6.6.10, "Example: SEM
Measurement", on page 587.
Manual operation: See " Abs Limit Start / Stop <n> " on page 305