User Manual

Remote Commands
R&S
®
FPL1000
574User Manual 1178.3370.02 ─ 03
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the SEM range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Relative limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: -50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
For a detailed example see Chapter 8.6.6.10, "Example: SEM
Measurement", on page 587.
Manual operation: See " Rel Limit Start / Stop <n> " on page 305
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STOP:ABS
<Limit>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see "Relative limit line functions" on page 295.
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the SEM range.
<PClass> 1..4
Power class for which the limit is defined.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<range>:
LIMit<PClass>:RELative:STOP:FUNCtion
on page 575).
Range: -400 to 400
*RST: -13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
For a detailed example see Chapter 8.6.6.10, "Example: SEM
Measurement", on page 587.
Manual operation: See " Rel Limit Start / Stop <n> " on page 305
Configuring and Performing Measurements