Parameters:
<State> ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
RELative
Manual operation: See " Limit Check <n> " on page 305
[SENSe:]ESPectrum<sb>:RANGe<range>:POINts:MINimum[:VALue]
<SweepPoint>
Defines the minimum number of sweep points for the range.
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the measurement range.
Parameters:
<SweepPoints> Minimum number of sweep points per range
Range: 1 to 32001
*RST: 1
Example:
SENSe1:ESPectrum:RANGe3:POINts:MIN:VALue 400
Manual operation: See " Min Sweep Points " on page 306
[SENSe:]ESPectrum<sb>:RANGe<range>:MLCalc <Function>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details see "Limit calculation for individ-
ual ranges" on page 299.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Configuring and Performing Measurements