578User Manual 1178.3370.02 ─ 03
[SENSe:]ESPectrum<sb>:RANGe<range>:SWEep:TIME:AUTO <State>
This command turns automatic selection of the sweep time for a SEM range on and
off.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
<range> 1..30
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation: See " Sweep Time Mode " on page 304
[SENSe:]ESPectrum<sb>:RANGe<range>:TRANsducer <Transducer>
This command selects a transducer factor for a SEM range.
Note that
the transducer must cover at least the span of the range
the x-axis has to be linear
the unit has to be dB
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range> 1..30
Selects the measurement range.
Parameters:
<Transducer> String containing the transducer file name, including the path
information.
Manual operation: See " Transducer Factor " on page 305
8.6.6.5 Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
[SENSe:]ESPectrum<sb>:BWID......................................................................................579
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa................................................................579
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe].............................................................. 579
[SENSe:]ESPectrum<sb>:RRANge?................................................................................579
[SENSe:]ESPectrum<sb>:RTYPe....................................................................................580
Configuring and Performing Measurements