Return values:
<RefRange> Number of the current reference range.
Range: 1 to 30
Usage: Query only
[SENSe:]ESPectrum<sb>:RTYPe <Type>
This command defines the type of the power reference.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
Parameters:
<Type> PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST: CPOWer
Manual operation: See " Power Reference Type " on page 309
8.6.6.6 Configuring the Power Classes
The following commands define the power classes for SEM measurements.
CALCulate<n>:LIMit<k>:ESPectrum<sb>:LIMits................................................................580
CALCulate<n>:LIMit<k>:ESPectrum<sb>:MODE...............................................................581
CALCulate<n>:LIMit<k>:ESPectrum<sb>:VALue...............................................................582
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass:COUNt..................................................582
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>[:EXCLusive]............................ 583
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:LIMit[:STATe].......................... 583
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:MAXimum............................... 584
CALCulate<n>:LIMit<k>:ESPectrum<sb>:PCLass<PClass>:MINimum................................ 584
CALCulate<n>:LIMit<k>:ESPectrum<sb>:LIMits <Max1>,<Max2>,<Max3>
This command sets or queries up to 4 power classes in one step. You can only define
values for the number of power classes defined by CALCulate<n>:LIMit<k>:
ESPectrum<sb>:PCLass:COUNt on page 582.
Suffix:
<n>, <k>
.
irrelevant
<sb> 1 | 2 | 3