User Manual
Remote Commands
R&S
®
FPL1000
586User Manual 1178.3370.02 ─ 03
Example:
CALC:ESP:PSE:AUTO OFF
Deactivates the list evaluation.
Manual operation: See " List Evaluation State (Result Summary)" on page 314
See " List Evaluation State " on page 336
CALCulate<n>:ESPectrum:PSEarch[:IMMediate]
CALCulate<n>:ESPectrum:PEAKsearch[:IMMediate]
This command initiates a list evaluation.
Suffix:
<n>
.
Window
<m> Marker
Usage: Event
CALCulate<n>:ESPectrum:PSEarch:MARGin <Threshold>
CALCulate<n>:ESPectrum:PEAKsearch:MARGin <Threshold>
This command defines the threshold of the list evaluation.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Margin> Range: -200 to 200
*RST: 200
Default unit: dB
Example:
CALC:ESP:PSE:MARG 100
Sets the margin to 100 dB.
Manual operation: See " Margin " on page 314
See " Margin " on page 336
CALCulate<n>:ESPectrum:PSEarch:PSHow <State>
CALCulate<n>:ESPectrum:PEAKsearch:PSHow <State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Configuring and Performing Measurements