User Manual

Remote Commands
R&S
®
FPL1000
593User Manual 1178.3370.02 ─ 03
[SENSe:]LIST:RANGe<range>:DETector <Detector>
This command selects the detector for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Detector> APEak
Autopeak
NEGative
minimum peak detector
POSitive
peak detector
SAMPle
sample detector
RMS
RMS detector
AVERage
average detector
*RST:
RMS
Manual operation: See " Detector " on page 333
[SENSe:]LIST:RANGe<range>:FILTer:TYPE <FilterType>
This command selects the filter type for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Manual operation: See " Filter Type " on page 332
[SENSe:]LIST:RANGe<range>[:FREQuency]:STARt <Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
smaller than the span the sweep list covers, the R&S FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
greater than the span the sweep list covers, the R&S FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
For more information see Chapter 7.2.7, "Spurious Emissions Measurement",
on page 327.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements