User Manual
Remote Commands
R&S
®
FPL1000
595User Manual 1178.3370.02 ─ 03
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation:AUTO <State>
This command turns automatic selection of the input attenuation for a spurious emis-
sion measurement range on and off.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Manual operation: See " RF Attenuation Mode " on page 334
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Preamp " on page 334
[SENSe:]LIST:RANGe<range>:INPut:GAIN[:VALue] <Gain>
This command selects the preamplification level for the range.
Suffix:
<range>
.
1..30
Selects the measurement range.
[SENSe:]LIST:RANGe<range>:LIMit:STARt <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Manual operation: See " Abs Limit Start / Abs Limit Stop " on page 335
Configuring and Performing Measurements