User Manual

Remote Commands
R&S
®
FPL1000
596User Manual 1178.3370.02 ─ 03
[SENSe:]LIST:RANGe<range>:LIMit:STATe
This command turns the limit check for all spurious emission measurement ranges on
and off (<range> is irrelevant).
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Limit Check " on page 334
[SENSe:]LIST:RANGe<range>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Level> Absolute limit at the stop frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Manual operation: See " Abs Limit Start / Abs Limit Stop " on page 335
[SENSe:]LIST:RANGe<range>:POINts[:VALue] <Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Points> For more information on sweep points see Chapter 7.6.1.8,
"How Much Data is Measured: Sweep Points and Sweep
Count", on page 395.
*RST: 1001
Manual operation: See "Sweep Points" on page 334
[SENSe:]LIST:RANGe<range>:RLEVel <RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<range>
.
1..30
Selects the measurement range.