Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
8.6.9.2 Configuring the Measurement
The following commands control the harmonic distortion measurement.
Useful commands for harmonic distortion measurements described elsewhere
●
CALCulate<n>:MARKer<m>:FUNCtion:CENTer on page 641
●
[SENSe:]SWEep:TIME:AUTO on page 656
Remote commands exclusive to harmonic distortion measurements
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO...............................613
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics......................................613
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet..............................................614
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO <State>
This command selects the resolution bandwidth of the harmonic in respect to the band-
width of the first harmonic.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
identical
ON | 1
a multiple
*RST: 1
Manual operation: See " Harmonic RBW Auto " on page 349
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics <NoHarmonics>
This command selects the number of harmonics that the R&S FPL1000 looks for.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<NoHarmonics> Range: 1 to 26
*RST: 10
Manual operation: See " Number of Harmonics " on page 349