8.6.10 Measuring the Third Order Intercept Point
● Determining the TOI..............................................................................................616
● Programming Example: Measuring the TOI..........................................................617
8.6.10.1 Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
●
CALCulate<n>:DELTamarker<m>:X on page 712
●
CALCulate<n>:DELTamarker<m>:X:RELative? on page 726
●
CALCulate<n>:DELTamarker<m>:Y? on page 727
●
CALCulate<n>:MARKer<m>:X on page 714
●
CALCulate<n>:MARKer<m>:Y? on page 727
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]......................................................... 616
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE...................................... 617
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?......................................................... 617
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe] <State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
CALCulate<n>:DELTamarker<m>:X command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
Configuring and Performing Measurements