User Manual

Remote Commands
R&S
®
FPL1000
617User Manual 1178.3370.02 ─ 03
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
This command initiates a search for signals in the current trace to determine the third
intercept point.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
ONCE
Example:
CALC:MARK:FUNC:TOI:SEAR ONCE
Executes the search for 2 signals and their intermodulation prod-
uct at the currently available trace.
Usage: Event
Manual operation: See " Search Signals " on page 357
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
This command queries the results for the third order intercept point measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 536.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<TOI> Third order intercept point.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:TOI ON
Switches the intercept measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:TOI:RES?
Outputs the measured value.
Usage: Query only
8.6.10.2 Programming Example: Measuring the TOI
This example demonstrates how to determine the TOI in a remote environment.
//-----------Configuring the measurement ------------
*RST
//Reset the instrument
Configuring and Performing Measurements