[SENSe:]LIST:POWer[:SEQuence]
This command configures and initiates the List Evaluation measurement.
The list can contain up to 200 entries (frequencies). You can define a different instru-
ment setup for each frequency that is in the list.
If you synchronize the measurement with *OPC, the R&S FPL1000 produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see [SENSe:]LIST:POWer:RESult?.
Parameters:
<Frequency> Defines the frequency. Each frequency corresponds to one list
entry.
Range: 0 to Fmax
Default unit: Hz
<RefLevel> Defines the reference level for a list entry.
Range: -130 to 30
Increment: 0.01
Default unit: dBm
<RFAttenuation> Defines the RF attenuation for a list entry.
Range: 0 to 70
Increment: 1
Default unit: dB
<FilterType> Selects the filter type for a list entry. For more information see
[SENSe:]BANDwidth[:RESolution]:TYPE.
<RBW> Defines the resolution bandwidth for a list entry.
<VBW> Defines the video bandwidth for a list entry.
<MeasTime> Defines the measurement time for a list entry.
Range: 1 µs to 16000 s
<TriggerLevel> The trigger level must be 0.
Return values:
<PowerLevel> Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with [SENSe:]LIST:POWer:SET.
The result is a list of floating point values separated by commas.
The unit depends on CALCulate<n>:UNIT:POWer.