[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
Center frequency
Reference level
Example:
ADJ:ALL
Usage: Event
Manual operation: See " Adjusting all Determinable Settings Automatically ( Auto
All )" on page 419
[SENSe:]ADJust:CONFigure[:LEVel]:DURation <Duration>
In order to determine the ideal reference level, the R&S FPL1000 performs a measure-
ment on the current input data. This command defines the length of the measurement
if [SENSe:]ADJust:CONFigure[:LEVel]:DURation:MODE is set to MANual.
Parameters:
<Duration> Numeric value in seconds
Range: 0.001 to 16000.0
*RST: 0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation: See " Changing the Automatic Measurement Time ( Meastime
Manual )" on page 420
[SENSe:]ADJust:CONFigure[:LEVel]:DURation:MODE <Mode>
In order to determine the ideal reference level, the R&S FPL1000 performs a measure-
ment on the current input data. This command selects the way the R&S FPL1000
determines the length of the measurement .
Parameters:
<Mode> AUTO
The R&S FPL1000 determines the measurement length auto-
matically according to the current input data.
MANual
The R&S FPL1000 uses the measurement length defined by
[SENSe:]ADJust:CONFigure[:LEVel]:DURation
on page 646.
*RST: AUTO
Setting Basic Spectrum RF Measurement Parameters