[SENSe:]BWIDth:VIDeo <Bandwidth>
[SENSe:]BANDwidth:VIDeo <Bandwidth>
This command defines the video bandwidth.
The command decouples the video bandwidth from the resolution bandwidths.
Parameters:
<Bandwidth> refer to data sheet
*RST: AUTO is set to ON
Example:
BAND:VID 10 kHz
Manual operation: See " VBW " on page 397
[SENSe:]BWIDth:VIDeo:AUTO <State>
[SENSe:]BANDwidth:VIDeo:AUTO <State>
This command couples and decouples the video bandwidth to the resolution band-
width.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
BAND:VID:AUTO OFF
Manual operation: See " VBW " on page 397
See " RBW/VBW " on page 398
See " Default Coupling " on page 399
[SENSe:]BWIDth:VIDeo:RATio <Ratio>
[SENSe:]BANDwidth:VIDeo:RATio <Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (RBW/VBW).
Parameters:
<Ratio> Range: 0,001 to 1000
*RST: 1
Example:
BAND:VID:RAT 3
Sets the coupling of video bandwidth to video bandwidth =
3*resolution bandwidth
Manual operation: See " RBW/VBW " on page 398
[SENSe:]BWIDth:VIDeo:TYPE <Mode>
[SENSe:]BANDwidth:VIDeo:TYPE <Mode>
This command selects the position of the video filter in the signal path.
Setting Basic Spectrum RF Measurement Parameters