Parameters:
<SweepCount> When you set a sweep count of 0 or 1, the R&S FPL1000 per-
forms one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation: See " Sweep/Average Count " on page 399
[SENSe:]SWEep:DURation? <Time>
This command provides an estimation of the total time required to capture the data and
process it. This time span may be considerably longer than the actual sweep time (see
[SENSe:]SWEep:TIME on page 656).
Tip: To determine the necessary timeout for data capturing in a remote control pro-
gram, double the estimated time and add 1 second.
Return values:
<Time>
Example:
SWE:TIME 1s
SWE:DUR?
Reply:
27.9734842578
Usage: Query only
Manual operation: See " Sweep Time " on page 398
See "Data capturing takes too long" on page 861
[SENSe:]SWEep:OPTimize <Mode>
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called partial span. The
partial span cannot be defined directly, but it can be optimized according to measure-
ment requirements.