User Manual

Remote Commands
R&S
®
FPL1000
656User Manual 1178.3370.02 ─ 03
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths. Note that this command queries only the time
required to capture the data, not to process it. To obtain an estimation of the total cap-
ture and processing time, use the [SENSe:]SWEep:DURation? command.
Parameters:
<Time> refer to data sheet
*RST: depends on current settings (determined automati-
cally)
Example:
SWE:TIME 10s
Manual operation: See " Sweep Time " on page 266
See " Sweep Time " on page 398
See " Sweep Time " on page 410
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation: See " Harmonic Sweep Time " on page 349
See " Sweep Time " on page 398
See " Default Coupling " on page 399
[SENSe:]SWEep:TYPE:USED
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type> SWE
Normal sweep
FFT
FFT mode
8.8.3 Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
Setting Basic Spectrum RF Measurement Parameters