Parameters:
<Unit> DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST: dBm
Example:
UNIT:POW DBM
Sets the power unit to dBm.
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel <ReferenceLevel>
This command defines the reference level (for all traces in all windows).
With a reference level offset ≠ 0, the value range of the reference level is modified by
the offset.
Suffix:
<n>, <t>
.
irrelevant
Parameters:
<ReferenceLevel> The unit is variable.
Range: see datasheet
*RST: 0 dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation: See " Reference Level " on page 385
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet <Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>, <t>
.
irrelevant
Parameters:
<Offset> Range: -200 dB to 200 dB
*RST: 0dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation: See " Shifting the Display ( Offset )" on page 385
8.8.3.2 Configuring the Attenuation
INPut:ATTenuation.........................................................................................................658
INPut:ATTenuation:AUTO...............................................................................................659
INPut:EATT:AUTO......................................................................................................... 659
INPut:ATTenuation <Attenuation>
This command defines the total attenuation for RF input.
Setting Basic Spectrum RF Measurement Parameters