INPut:GAIN:STATe <State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
The input signal is amplified by 20 dB if the preamplifier option is activated.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
INP:GAIN:STAT ON
Switches on 20 dB preamplification.
Manual operation: See " Preamplifier " on page 387
8.8.3.4 Scaling the Y-Axis
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]................................................................... 660
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE............................................... 660
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE.........................................................661
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision.................................................... 661
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition................................................... 661
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing.................................................................. 662
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe] <Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Range> Range: 1 dB to 200 dB
*RST: 100 dB
Example:
DISP:TRAC:Y 110dB
Manual operation: See " Range " on page 388
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
Window
<t> irrelevant