User Manual
Remote Commands
R&S
®
FPL1000
661User Manual 1178.3370.02 ─ 03
Manual operation: See " Auto Scale Once " on page 388
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE <Mode>
This command selects the type of scaling of the y-axis (for all traces).
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Mode> ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST: ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation: See " Scaling " on page 389
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision <Value>
This remote command determines the grid spacing on the Y-axis for all diagrams,
where possible.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Value> numeric value WITHOUT UNIT (unit according to the result dis-
play)
Defines the range per division (total range = 10*<Value>)
*RST: depends on the result display
Example:
DISP:TRAC:Y:PDIV 10
Sets the grid spacing to 10 units (e.g. dB) per division
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition <Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S FPL1000 adjusts the scaling of the y-axis accordingly.
Suffix:
<n>
.
Window
Setting Basic Spectrum RF Measurement Parameters