User Manual

Remote Commands
R&S
®
FPL1000
662User Manual 1178.3370.02 ─ 03
<t> irrelevant
Parameters:
<Position> 0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST: 100 PCT = frequency display; 50 PCT = time dis-
play
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation: See " Ref Level Position " on page 388
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing <ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Manual operation: See " Scaling " on page 389
8.8.4 Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here. The tasks for manual operation are described in
Chapter 7.7, "Trigger and Gate Configuration", on page 405.
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the Triggering Conditions...................................................................663
Configuring Gated Measurements........................................................................ 666
Setting Basic Spectrum RF Measurement Parameters