662User Manual 1178.3370.02 ─ 03
<t> irrelevant
Parameters:
<Position> 0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST: 100 PCT = frequency display; 50 PCT = time dis-
play
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation: See " Ref Level Position " on page 388
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing <ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Manual operation: See " Scaling " on page 389
8.8.4 Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here. The tasks for manual operation are described in
Chapter 7.7, "Trigger and Gate Configuration", on page 405.
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
● Configuring the Triggering Conditions...................................................................663
● Configuring Gated Measurements........................................................................ 666
Setting Basic Spectrum RF Measurement Parameters