8.8.4.1 Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
TRIGger[:SEQuence]:DTIMe...........................................................................................663
TRIGger[:SEQuence]:HOLDoff[:TIME]..............................................................................663
TRIGger[:SEQuence]:IFPower:HOLDoff...........................................................................663
TRIGger[:SEQuence]:IFPower:HYSTeresis...................................................................... 664
TRIGger[:SEQuence]:LEVel[:EXTernal<port>].................................................................. 664
TRIGger[:SEQuence]:LEVel:IFPower...............................................................................664
TRIGger[:SEQuence]:LEVel:IQPower.............................................................................. 665
TRIGger[:SEQuence]:SLOPe.......................................................................................... 665
TRIGger[:SEQuence]:SOURce........................................................................................665
TRIGger[:SEQuence]:TIME:RINTerval............................................................................. 666
TRIGger[:SEQuence]:DTIMe <DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime> Dropout time of the trigger.
Range: 0 s to 10.0 s
*RST: 0 s
Manual operation: See " Drop-Out Time " on page 412
TRIGger[:SEQuence]:HOLDoff[:TIME] <Offset>
Defines the time offset between the trigger event and the start of the sweep.
A negative offset is possible for time domain measurements.
For the trigger sources "External" or "IF Power", a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger offset as
well.
Parameters:
<Offset> For measurements in the frequency domain, the range is 0 s to
30 s.
For measurements in the time domain, the range is the negative
sweep time to 30 s.
*RST: 0 s
Example:
TRIG:HOLD 500us
Manual operation: See " Trigger Offset " on page 412
TRIGger[:SEQuence]:IFPower:HOLDoff <Period>
This command defines the holding time before the next trigger event.
Setting Basic Spectrum RF Measurement Parameters