User Manual

Remote Commands
R&S
®
FPL1000
669User Manual 1178.3370.02 ─ 03
Parameters:
<Type> LEVel
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open
until the signal disappears.
Note: If you perform gated measurements in combination with
the IF Power trigger, the R&S FPL1000 ignores the holding time
for frequency sweep, FFT sweep, zero span and I/Q mode mea-
surements.
EDGE
The trigger event for the gate to open is the detection of the sig-
nal edge.
After the gate signal has been detected, the gate remains open
until the gate length is over.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation: See " Gate Mode " on page 417
8.8.5 Configuring the Data Input and Output
The following commands are required to configure data input and output.
RF Input................................................................................................................ 669
Working with Power Sensors................................................................................ 671
Configuring the Outputs........................................................................................ 681
8.8.5.1 RF Input
INPut:ATTenuation:PROTection:RESet............................................................................669
INPut:IMPedance...........................................................................................................670
INPut:SELect.................................................................................................................670
INPut:UPORt:STATe......................................................................................................670
INPut:UPORt[:VALue]?...................................................................................................670
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occurred and the protection mechanism intervened. The
error status bit (bit 3 in the STAT:QUES:POW status register) and the INPUT OVLD
message in the status bar are cleared.
(See STATus:QUEStionable:POWer[:EVENt]? on page 842 and "STA-
Tus:QUEStionable:POWer Register" on page 194).
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see "RF Input Protection" on page 364.
Usage:
Event
Setting Basic Spectrum RF Measurement Parameters