User Manual
Remote Commands
R&S
®
FPL1000
671User Manual 1178.3370.02 ─ 03
Example:
INP:UPOR?
//Result: #B00100100
Pins 5 and 7 are active.
Usage: Query only
8.8.5.2 Working with Power Sensors
The following commands describe how to work with power sensors.
These commands require the use of a Rohde & Schwarz power sensor. For a list of
supported sensors, see the data sheet.
● Configuring Power Sensors.................................................................................. 671
● Configuring Power Sensor Measurements........................................................... 672
● Triggering with Power Sensors............................................................................. 679
Configuring Power Sensors
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]......................... 671
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?..........................................................671
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine......................................................672
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe] <State>
This command turns automatic assignment of a power sensor to the power sensor
index on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SYST:COMM:RDEV:PMET:CONF:AUTO OFF
Manual operation: See " Select " on page 368
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
This command queries the number of power sensors currently connected to the
R&S FPL1000.
Parameters:
<NumberSensors> Number of connected power sensors.
Example:
SYST:COMM:RDEV:PMET:COUN?
Usage: Query only
Manual operation: See " Select " on page 368
Setting Basic Spectrum RF Measurement Parameters