<p> 1...4
Power sensor index
Parameters:
ONCE
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Usage: Event
Manual operation: See "Setting the Reference Level from the Measurement Meas -
> Ref " on page 369
CALCulate<n>:PMETer<p>:RELative:STATe
<State>
This command turns relative power sensor measurements on and off.
Suffix:
<n>
.
Window
<p> 1...4
Power sensor index
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:PMET2:REL:STAT ON
Activates the relative display of the measured value for power
sensor 2.
FETCh:PMETer<p>?
This command queries the results of power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Return values:
<Level> Power level that has been measured by a power sensor.
The unit is either dBm (absolute measurements) or dB (relative
measurements).
Usage: Query only
READ:PMETer<p>?
This command initiates a power sensor measurement and queries the results.