[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt <NumberReadings>
This command sets the number of power readings included in the averaging process of
power sensor measurements.
Extended averaging yields more stable results for power sensor measurements, espe-
cially for measurements on signals with a low power, because it minimizes the effects
of noise.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<NumberReadings> An average count of 0 or 1 performs one power reading.
Range: 0 to 256
Increment: binary steps (1, 2, 4, 8, ...)
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Manual operation: See " Average Count ( Number of Readings )" on page 370
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe] <State>
This command turns averaging for power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation: See " Meas Time/Average " on page 369
[SENSe:]PMETer<p>:ROFFset[:STATe] <State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index