User Manual

Remote Commands
R&S
®
FPL1000
682User Manual 1178.3370.02 ─ 03
Output functions require the option R&S FPL1-B5 to be installed on the R&S FPL1000.
DIAGnostic:SERVice:NSOurce........................................................................................682
OUTPut:IF[:SOURce]..................................................................................................... 682
OUTPut:IF:IFFRequency................................................................................................ 683
OUTPut:UPORt:STATe.................................................................................................. 683
OUTPut:UPORt[:VALue].................................................................................................683
OUTPut:UPORt:WTRigger:POLarity................................................................................ 684
SYSTem:SPEaker[:STATe].............................................................................................684
SYSTem:SPEaker:VOLume............................................................................................684
DIAGnostic:SERVice:NSOurce <State>
This command turns the 28 V supply of the BNC connector labeled NOISE SOURCE
CONTROL on the R&S FPL on and off.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
DIAG:SERV:NSO ON
Manual operation: See "Noise Source Control" on page 374
OUTPut:IF[:SOURce] <Source>
Defines the type of signal available at the IF/VIDEO connector of the R&S FPL.
Parameters:
<Source> IF
The measured IF value is available at the IF/VIDEO output con-
nector.
The frequency at which the IF value is provided is defined using
the OUTPut:IF:IFFRequency command.
VIDeo
The displayed video signal (i.e. the filtered and detected IF sig-
nal, 200mV) is available at the IF/VIDEO output connector.
This setting is required to provide demodulated audio frequen-
cies at the output.
*RST: IF
Example:
OUTP:IF VID
Selects the video signal for the IF/VIDEO output connector.
Manual operation: See "Data Output" on page 374
Setting Basic Spectrum RF Measurement Parameters