User Manual

Remote Commands
R&S
®
FPL1000
683User Manual 1178.3370.02 ─ 03
OUTPut:IF:IFFRequency <Frequency>
This command defines the frequency for the IF output of the R&S FPL. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF/VIDEO output is configured
for IF.
Parameters:
<Frequency> *RST: 50.0 MHz
Manual operation: See "Data Output" on page 374
OUTPut:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the AUX PORT connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S FPL1000.
Parameters:
<State> ON | 1
User port is switched to OUTPut
OFF | 0
User port is switched to INPut
*RST: 0
OUTPut:UPORt[:VALue] <Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit 7 6 5 4 3 2 1 0
Pin N/A N/A 5 3 4 7 6 2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern. If the user port is programmed
to input instead of output (see INPut:UPORt:STATe on page 670), the output value is
temporarily stored.
Parameters:
<Value> bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range: #B00000000 to #B00111111
Example:
OUTP:UPOR #B00100100
Sets pins 5 and 7 to 5 V.