User Manual
Remote Commands
R&S
®
FPL1000
685User Manual 1178.3370.02 ─ 03
Parameters:
<Volume> Percentage of the maximum possible volume.
Range: 0 to 1
*RST: 0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
Manual operation: See "Data Output" on page 374
See "Audio Output Volume" on page 453
8.8.6 Zooming into the Display
8.8.6.1 Using the Single Zoom
DISPlay[:WINDow<n>]:ZOOM:AREA............................................................................... 685
DISPlay[:WINDow<n>]:ZOOM[:STATe]............................................................................ 686
DISPlay[:WINDow<n>]:ZOOM:AREA <x1>,<y1>,<x2>,<y2>
This command defines the zoom area.
To define a zoom area, you first have to turn the zoom on.
1
= origin of coordinate system (x1 = 0, y1 = 0)
2 = end point of system (x2 = 100, y2= 100)
3 = zoom area (e.g. x1 = 60, y1 = 30, x2 = 80, y2 = 75)
Suffix:
<n>
.
Window
Parameters:
<x1>,<y1>,
<x2>,<y2>
Diagram coordinates in % of the complete diagram that define
the zoom area.
The lower left corner is the origin of coordinate system. The
upper right corner is the end point of the system.
Range: 0 to 100
Default unit: PCT
Manual operation: See " Single Zoom " on page 509
Setting Basic Spectrum RF Measurement Parameters