User Manual
Remote Commands
R&S
®
FPL1000
690User Manual 1178.3370.02 ─ 03
Manual operation: See " Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6 "
on page 467
See " Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)"
on page 470
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing:APERture <Percentage>
This command defines the degree (aperture) of the trace smoothing, if DISPlay[:
WINDow<n>]:TRACe<t>:SMOothing[:STATe] TRUE.
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<Percentage> Range: 1 to 50
*RST: 2
Default unit: PCT
Example:
DISP3:TRAC2:SMO:APER 5
Defines an aperture of 5% for trace 2 in window 3
Manual operation: See " Smoothing " on page 468
DISPlay[:WINDow<n>]:TRACe<t>:SMOothing[:STATe] <State>
This command turns trace smoothing for a particular trace on and off.
If enabled, the trace is smoothed by the value specified using DISPlay[:
WINDow<n>]:TRACe<t>:SMOothing:APERture on page 690.
For more information see "Trace Smoothing" on page 465.
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
DISP3:TRAC2:SMO ON
Turns on trace smoothing for trace 2 in window 3
Manual operation: See " Smoothing " on page 468
[SENSe:]AVERage<n>:COUNt <AverageCount>
This command defines the number of sweeps that the application uses to average
traces.
Setting Basic Spectrum RF Measurement Parameters