User Manual

Remote Commands
R&S
®
FPL1000
702User Manual 1178.3370.02 ─ 03
CALCulate<n>:MATH:STATe <State>
This command turns the trace mathematics on and off.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:MATH:STAT ON
Turns on trace mathematics.
Manual operation: See " Trace Math Function " on page 487
See " Trace Math Off " on page 488
8.8.7.4 Retrieving Trace Results
This chapter describes how to retrieve data from standard traces.
For spectrograms see also Chapter 8.8.8.6, "Marker Search (Spectrograms)",
on page 728.
For details on the format of the retrieved trace data see also Chapter 8.8.7.5, "Formats
for Returned Values: ASCII Format and Binary Format", on page 705.
Commands required to export traces (and other result data) are described in Chap-
ter 8.9.5, "Storing Measurement Results", on page 810.
FORMat[:DATA].............................................................................................................702
TRACe<n>[:DATA].........................................................................................................703
TRACe<n>[:DATA]:MEMory?..........................................................................................704
TRACe<n>[:DATA]:X?....................................................................................................705
FORMat[:DATA] <Format>
This command selects the data format that is used for transmission of trace data from
the R&S FPL1000 to the controlling computer.
Note that the command has no effect for data that you send to the R&S FPL1000. The
R&S FPL1000 automatically recognizes the data it receives, regardless of the format.
For details on data formats see Chapter 8.8.7.5, "Formats for Returned Values: ASCII
Format and Binary Format", on page 705.
Setting Basic Spectrum RF Measurement Parameters