User Manual
Remote Commands
R&S
®
FPL1000
703User Manual 1178.3370.02 ─ 03
Parameters:
<Format> ASCii
ASCii format, separated by commas.
This format is almost always suitable, regardless of the actual
data format. However, the data is not as compact as other for-
mats may be.
REAL,32
32-bit floating-point numbers (according to IEEE 754) in the
"definite length block format".
In the Spectrum application, the format setting REAL is used for
the binary transmission of trace data.
For I/Q data, 8 bytes per sample are returned for this format set-
ting.
*RST:
ASCII
Example:
FORM REAL,32
TRACe<n>[:DATA]
<Trace>,<Data>
TRACe<n>[:DATA]? <ResultType>
This command queries current trace data and measurement results.
The data format depends on FORMat[:DATA].
Suffix:
<n>
.
Window
Parameters:
<Trace> Selects the trace to write the data to.
TRACE1 | ... | TRACE6
<Data> Contains the data to transfer.
Query parameters:
<ResultType> Selects the type of result to be returned.
TRACE1 | ... | TRACE6
Returns the trace data for the corresponding trace.
For details see Table 8-5.
LIST
Returns the results of the peak list evaluation for Spurious Emis-
sion and Spectrum Emission Mask measurements.
For SEM measurements, one peak per range is returned.
For details see Table 8-6.
SPURious
Returns the peak list of Spurious Emission measurements.
SPECtrogram | SGRam
Returns the results of the spectrogram result display.
For details see Table 8-7.
Return values:
<TraceData> For more information see tables below.
Setting Basic Spectrum RF Measurement Parameters