715User Manual 1178.3370.02 ─ 03
Parameters:
<DisplayMode> ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST: AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation: See " Marker Table Display " on page 428
DISPlay:MINFo[:STATe] <DisplayMode>
This command turns the marker information in all diagrams on and off.
Parameters:
<DisplayMode> ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST: 1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation: See " Marker Info " on page 429
CALCulate<n>:MARKer<m>:X:SSIZe <StepSize>
This command selects the marker step size mode for all markers in all windows.
The step size defines the distance the marker moves when you move it with the rotary
knob.
It therefore takes effect in manual operation only.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<StepSize> STANdard
the marker moves from one pixel to the next
POINts
the marker moves from one sweep point to the next
*RST: POINts
Example:
CALC:MARK:X:SSIZ STAN
Sets the marker step size to one pixel.
Manual operation: See " Marker Stepsize " on page 429
Setting Basic Spectrum RF Measurement Parameters