User Manual

Remote Commands
R&S
®
FPL1000
718User Manual 1178.3370.02 ─ 03
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Limit> The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST: right diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation: See " Search Limits ( Left / Right )" on page 285
See " Left Limit / Right Limit " on page 342
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe] <State>
This command adjusts the marker search range to the zoom area for all markers in all
windows.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:MARK:X:SLIM:ZOOM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation: See " Use Zoom Limits " on page 432
CALCulate<n>:THReshold <Level>
This command defines a threshold level for the marker peak search (for all markers in
all windows).
Suffix:
<n>
.
irrelevant
Parameters:
<Level> Numeric value. The value range and unit are variable.
*RST: -120 dBm
Example:
CALC:THR -82DBM
Sets the threshold value to -82 dBm.
Setting Basic Spectrum RF Measurement Parameters