User Manual

Remote Commands
R&S
®
FPL1000
738User Manual 1178.3370.02 ─ 03
CALCulate<n>:DELTamarker<m>:SGRam:Y:MINimum[:PEAK]
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum[:PEAK]
This command moves a delta marker vertically to the minimum level for the current fre-
quency.
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command first looks for the peak level in the
whole spectrogram and moves the marker vertically to the minimum level.
Suffix:
<n>
.
Window
<m> Marker
Usage: Event
8.8.8.7 Fixed Reference Marker Settings
The following commands configure a fixed reference marker.
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK].................... 738
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X............................................ 738
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y............................................ 739
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet................................ 739
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe]..............................................739
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
This command moves the fixed reference marker to the peak power.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:DELT:FUNC:FIX:RPO:MAX
Sets the reference point level for delta markers to the peak of
the selected trace.
Usage: Event
Manual operation: See "Defining a Fixed Reference" on page 429
See " Defining Reference Point " on page 445
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X <RefPoint>
This command defines the horizontal position of the fixed delta marker reference point.
The coordinates of the reference may be anywhere in the diagram.
Suffix:
<n>
.
Window
<m> Marker
Setting Basic Spectrum RF Measurement Parameters