User Manual

Remote Commands
R&S
®
FPL1000
739User Manual 1178.3370.02 ─ 03
Parameters:
<RefPoint> Numeric value that defines the horizontal position of the refer-
ence.
For frequency domain measurements, it is a frequency in Hz.
For time domain measurements, it is a point in time in s.
*RST: Fixed Reference: OFF
Example:
CALC:DELT:FUNC:FIX:RPO:X 128 MHz
Sets the frequency reference to 128 MHz.
Manual operation: See "Defining a Fixed Reference" on page 429
See " Defining Reference Point " on page 445
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y
<RefPointLevel>
This command defines the vertical position of the fixed delta marker reference point.
The coordinates of the reference may be anywhere in the diagram.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<RefPoint> Numeric value that defines the vertical position of the reference.
The unit and value range is variable.
*RST: Fixed Reference: OFF
Example:
CALC:DELT:FUNC:FIX:RPO:Y -10dBm
Sets the reference point level for delta markers to -10 dBm.
Manual operation: See "Defining a Fixed Reference" on page 429
See " Defining Reference Point " on page 445
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet <Offset>
This command defines a level offset for the fixed delta marker reference point.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Offset> Numeric value
*RST: 0
Default unit: dB
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe] <State>
This command activates or deactivates a marker that defines a fixed reference point
for relative marker analysis.
Setting Basic Spectrum RF Measurement Parameters